Ssb phase noise – VXI 3002 User Manual

Page 84

Advertising
background image

5-9

SSB phase noise

Test procedure

UUT

C2563

RF

LO

RF OUTPUT

Signal generator

RF OUT

Measuring

receiver

Fig. 5-6 SSB phase noise test set-up

(1) Connect the test equipment as shown in Fig. 5-6.

(2) On the UUT set:

Carr Freq

470 MHz

RF Level

0 dBm

(3) On the measuring receiver:

Tune the receiver to 470 MHz.

Select 24.0 SPCL to enter selective power measurement mode.

Select 23.1 SPCL to set the LO to external.

(4) Set the signal generator to a carrier frequency of 470.455 MHz, RF level 0 dBm.

(5) On the measuring receiver:

Select 24.5 SPCL to establish the IF reference value (in volts).

Select 24.6 SPCL to set the reference to 0 dBm.

(6) Fine tune the signal generator frequency until a maximum value is displayed on the measuring

receiver.

(7) Offset the signal generator by 20 kHz.

(8) On the measuring receiver:

Select 24.7 SPCL to normalize the measurement for a 1 Hz bandwidth.

(9) Offset the signal generator by 20 kHz.

(10) Measure the level on the receiver (the SSB phase noise in a 1 Hz bandwidth) checking that

the result is within the specification shown in Table 5-21.

Advertising