3-22

Memory - triggering

For external triggering methods, refer to 'Trigger source' at the end of this section.

MTRIG

:ON

Enables memory recall triggering to be activated by *TRG command or
by external triggering

:OFF

Disable memory recall triggering

Data type : None

Allowed suffices : None

Default suffix : None

Examples: MTRIG:ON

MTRIG:OFF

MTRIG?

Prepares message containing information on memory triggering state in the
following format:

:MTRIG:<status>

where: <status> is a program mnemonic indicating whether the

memory recall triggering is enabled (:ON) or disabled (:OFF)

Example: :MTRIG:ON

Memory - protection

These commands enable you to either write protect a block of stores (or a single store) to prevent
accidental overwriting or to unprotect it. For a single store set both start and stop numbers the
same. Note that any protection applied to RAM will be lost once the instrument has been switched
off.

MPROT

[not used alone]

:START

Set the start of the memory block which is to be protected/unprotected

:STOP

Set the end of the memory block which is to be protected/unprotected

Data type : Decimal Numeric Program Data

Allowed suffices : None

Default suffix : None

:ON

Set memory protection ON for the selected memory block

:OFF

Set memory protection OFF (i.e. unprotected) for the selected memory
block

Data type : None

Allowed suffices : None

Default suffix : None

Examples: MPROT:START 100;STOP 150

MPROT:ON