5-16

(8) Set the UUT to RF level +19 dBm and repeat (2) to (5) using Table 5-33.

AM scale shape

Test procedure

(1) Connect the test equipment as shown in Fig. 5-7.

(2) On the UUT set:

Carr Freq

100 MHz

RF Level

0 dBm

Mod Mode

AM Internal

Source

On

Mod

On

AM1

Level

30%

(3) On the modulation meter, select CAL, AM, 300 Hz

Þ

3.4 kHz LF filter.

(4) Measure the AM accuracy at the depths shown in Table 5-34 checking that the results are

within specification.

External AM frequency response (ALC off, DC coupled)

Test procedure

100 Hz to 30 kHz

(1) Connect the test equipment as shown in Fig. 5-8.

(2) On the UUT set:

Carr Freq

400 MHz

RF

Level

−4 dBm

Mod Mode

AM External

AM1

Level

80%

Source

On

Mod

On

DC

coupling

(3) Set the function generator to give 1 V RMS, 1 kHz sine wave.

(4) On the modulation meter select CAL, AM, 10 Hz

Þ

300 kHz filter.

(5) On the modulation meter check that the AM reading is between 76% and 84%, then set a

reference using the relative function.

(6) Record the absolute reading for use in the formula in (16) below.

(7) Set the function generator to each of the frequencies shown in Table 5-35 checking that the

relative readings on the modulation meter are within specification.

(8) Set the UUT RF level to +7 dBm and repeat (3) to (7) using Table 5-36.

(9) Set the UUT RF level to +13 dBm and repeat (3) to (7) using Table 5-37.