Memory stores

Carrier store

The non-volatile carrier frequency store has 100 locations numbered 0 to 99 for the storage of
carrier frequency only. This store can be used to apply a set of test conditions to a range of
frequencies. For example, if you wish to use the same modulation at a variety of frequencies you
can use the carrier store to set the instrument to each of the frequencies in turn. When a carrier
store is recalled it will only replace the current carrier frequency - all the other settings will remain
unchanged.

Full store

The non-volatile full store has 100 locations numbered 100 to 199 for the storage of instrument
settings. This store is used to store those parameters which currently affect the RF output; carrier
frequency, RF level, modulations in use, on/off and source information and the two modulation
oscillator frequencies in use.

A full store contains the following information:

Carrier frequency setting

Carrier frequency step size

RF level setting

RF level step size

All modulation settings

All modulation step sizes

Modulation mode and status

The active modulation frequencies

The modulation frequency step size

All sweep settings

RAM store

The volatile RAM store has 100 locations numbered from 200 to 299 for the full storage of
instrument settings. The parameters stored are the same as those for the full store. However, the
RAM store has no long term wear-out mechanism and is therefore recommended for use in ATE
programs where all the settings to be used in a test sequence are initially declared and then recalled.
This results in a reduction of the 488.2 message overhead.

Memory - store

STO

Store 0-299 (short form)

:MEM

Store 0-299

:CFRQ

Carrier Freq Store 0-99

:FULL

Full Store 100-199

:RAM

RAM Store 200-299

Data type : Decimal Numeric Program Data

Allowed suffices : None

Default suffix : None

Examples: STO:FULL 112

STO:CFRQ 83