5-20

Pulse modulation rise and fall time

C2567

OUTPUT

Function generator

RF
OUT

UUT

PULSE

INPUT

Oscilloscope

Fig. 5-12 Pulse modulation rise and fall time test set-up

Test procedure

(1) Connect the test equipment as shown in Fig. 5-12.

(2) On the UUT set:

Carr Freq

50 MHz

RF Level

+7 dBm

Pulse

ON

(3) Set the function generator to produce 10 kHz, 0 V to +5 V square wave.

(4) Adjust the oscilloscope controls such that the rise time of the envelope can be measured.

(5) Measure the rise time between the 10% to 90% points checking that it is within the

specification shown in Table 5-42.

(6) Repeat (4) to (5) for the fall time of the envelope.

Modulation oscillator

Specification

Frequency range

0.01 Hz to 20 kHz.

Resolution

0.01 Hz for frequencies up to 100 Hz,
0.1 Hz for frequencies up to 1 kHz,
1 Hz for frequencies up to 20 kHz.

Frequency accuracy

As frequency standard.

Distortion

Less than 0.1% THD at 1 kHz.

Waveforms

Sine to 20 kHz, triangle or square wave to 3 kHz.

Audio output

The modulation oscillator signal is available on a front-panel BNC connector at a level of
2 V RMS EMF from a 600

Ω source impedance.

Test equipment

Description Minimum

specification Example

Frequency
counter

9 kHz to 2.4 GHz

Agilent 53181A
with option 030

Audio analyzer

Capable of measuring THD of 0.01% at 1 kHz

Rohde &
Schwarz UPA3