Fm scale shape – VXI 3002 User Manual

Page 86

Advertising
background image

5-11

(1) Connect the test equipment as shown in Fig. 5-7.

(2) On the UUT set:

Carr Freq

10 MHz

RF Level

0 dBm

FM1 Level

100 kHz

Mod

On

Source

On

(3) On the modulation meter select CAL, FM, 50 Hz

Þ 15 kHz filter.

(4) Measure the FM accuracy and distortion at the carrier frequencies shown in Table 5-22

checking that the results are within specification.

FM scale shape

Test procedure

(1) Connect the test equipment as shown in Fig. 5-7.

(2) On the UUT set:

Carr Freq

15 MHz

RF Level

0 dBm

FM1 Level

100 kHz

Mod

On

Source

On

(3) On the modulation meter select CAL, FM, 50 Hz

Þ 15 kHz filter.

(4) Measure the FM accuracy at the deviations shown in Table 5-23 checking that the results are

within specification.