Pulse modulation, Pulse modulation rf level frequency response – VXI 3002 User Manual

Page 93

Advertising
background image

5-18

Pulse modulation

Does not apply to instruments fitted with Option 11 — refer to Annex A.

Specification

Carrier frequency range

32 MHz to 2.4 GHz, usable to 10 MHz.

RF level range

Maximum guaranteed output is reduced to +20 dBm, +14 dBm above 1.2 GHz.

RF level accuracy

When pulse modulation is enabled, adds ±0.5 dB to the RF level accuracy.

ON/OFF ratio

Better than 45 dB below 1.2 GHz.
Better than 40 dB above 1.2 GHz.

Rise and fall time

Less than 10

µs.

Test equipment

Description Minimum

specification Example

Power meter

±0.1 dB from 9 kHz to 2.4 GHz

IFR 6960B and
6912

Spectrum analyzer

Frequency coverage 32 MHz to 2.4 GHz

Anritsu MS2602A

Oscilloscope

100 MHz bandwidth

Tektronix TDS 220

Function generator

DC to 10 kHz square wave

Agilent 3325B

Pulse modulation RF level frequency response

Fig. 5-10 Pulse modulation test set-up

Test procedure

(1) Perform AUTO ZERO and AUTO CAL on the power meter.

(2) Connect the test equipment as shown in Fig. 5-10.

(3) On the UUT set:

Carr Freq

32 MHz

RF

Level

−7 dBm

Pulse

ON

(4) Set the function generator to provide +5 V DC. The RF output will now be enabled.

(5) Record the output level measured by the power meter against each of the carrier frequencies

shown in Table 5-38 checking that the results are within specification.

(6) Set the UUT RF level to +4 dBm and repeat (5) using Table 5-39.