5-17

0 Hz (DC)

To measure the AM depth at DC, it will be necessary to use the DC offset facility on the function
generator proceeding as follows:

6960B

RF power meter

6912

Power

sensor

C2566

OUTPUT

Temporary
connection

SENSOR
INPUT

Function generator

DVM

RF
OUT

50
load

UUT

EXT MOD

INPUT

Fig. 5-9 0 Hz external AM and distortion test set-up

(9) Connect the test equipment as shown in Fig. 5-9.

(10) Set the function generator to +1.4142 V DC (temporarily connect the function generator

output to the DVM and set this voltage as close as possible to +1.4142 V).

(11) Measure the power on the power meter.


P1 _____

(12) Set the function generator to

−1.4142 V DC (temporarily connect the function generator

output to the DVM and set this voltage as close as possible to

−1.4142 V).

(13) Measure the power on the power meter.


P2 _____

(14) Subtract P2 from P1 (= x).

(15) Calculate the modulation depth using the formula:

пю

п

э

ь

по

п

н

м

(-x/20)

(-x/20)

10

+

1

10

-

1

=

AM(%)

(16) Calculate the 0 Hz response relative to 1 kHz using the following formula, recording the

result in Table 5-35:

ю

э

ь

о

н

м

)

15

(

in

recorded

Figure

)

6

(

in

recorded

Figure

log

20

10

(17) Set the UUT RF level to +7 dBm and repeat (10) to (16) using Table 5-36.

(18) Set the UUT RF level to +13 dBm and repeat (10) to (16) using Table 5-37.