Attenuator hold, Miscellaneous commands – VXI 3002 User Manual

Page 57

Advertising
background image

3-27

Miscellaneous commands

Attenuator hold

The ATTEN:LOCK command allows you to reduce the RF level by at least another 10 dB without
the step attenuator operating.

ATTEN

[not used alone]

:LOCK

Lock the Attenuators

:UNLOCK

Unlock the Attenuators

Data type : None

Allowed suffices : None

Default suffix : None

Example: ATTEN:LOCK

ATTEN?

Prepares message containing information on whether the Attenuators are
locked or unlocked in the following format:

:ATTEN:<status>

where

<status> is a program mnemonic indicating whether the
attenuators are locked or unlocked.

Example: :ATTEN:LOCK

Table 3-2 below applies to software versions 1.03 onwards. Maximum level with attenuator lock
on will reduce if AM or pulse modulation is applied. Any user programmed limits for RF level or
offsets will also affect the attenuator lock ranges.

Table 3-2 Attenuator lock ranges

RF level

setting (dBm)

Atten. lock

(min. dBm)

Atten. lock

(max. dBm)

from to

+26.0 +18.0 +7

+25

+17.9 +7.1 −4

+24

+7.0 +7.0 −4

+13

+6.9 -4.0

−15

+13

−4.1

−15.0

−26

+2

−15.1

−26.0

−37

−9

−26.1

−37.0

−48

−20

−37.1

−48.0

−59

−31

−48.1

−59.0

−70

−42

−59.1

−70.0

−81

−53

−70.1

−81.0

−82

−54

−81.1

−92.0

−103

−75

−92.1

−103.0

−114

−86

−103.1

−114.0

−125

−97

−114.1

−125.0

−136

−108

−125.1

−136.0

−137

−119

−136.1

−137.0

−137

−130