Modulation oscillator frequencies

Test procedure

C5359

INPUT

LF OUTPUT

UUT

Counter

Fig. 5-13 Modulation oscillator frequency test set-up

(1) Connect the test equipment as shown in Fig. 5-13.

(2) On the UUT set:

FM1 Freq

10 Hz

(3) Record the frequency measured by the counter against each of the modulation oscillator

frequencies shown in Table 5-43.

Modulation oscillator distortion and LF output flatness

Test procedure

LF OUTPUT

UUT

C2569

INPUT

Audio analyzer

Fig. 5-14 Modulation oscillator distortion test set-up

(1) Connect the test equipment as shown in Fig. 5-14.

(2) On the UUT set:

FM1 Freq

1 kHz