Memory-recall

There are three types of recall: carrier, full and RAM. Both carrier and full stores are non-volatile.
The contents of the RAM store are lost when the instrument is switched off.

Carrier recall

The non-volatile carrier frequency store has 100 locations numbered 0 to 99 for carrier frequency
only. These can be recalled and used in conjunction with full recall to apply a set of test conditions
to a range of frequencies.

Full recall

The non-volatile full store has 100 locations numbered 100 to 199 for the storage of instrument
settings. These stores may be recalled and used to reset the instrument's parameters to those which
affect the RF output: carrier frequency, RF level, modulations in use, on/off and source information
and the two modulation oscillator frequencies in use.

RAM recall

The volatile RAM store has locations numbered 200 to 299 for the full storage of instrument
settings. The parameters that are recalled are the same as those for full recall.

Recalling default settings

For a list of the default settings see Table 3-1.

Memory-recall

To recall the factory default settings, press the RCL 999.

RCL

Recall Store 0-299 (short form)

:MEM

Recall Store 0-299

:CFRQ

Recall Carrier Freq Store 0-99

:FULL

Recall Full Store 100-199

:RAM

Recall RAM Store 200-299

Data type : Decimal Numeric Program Data

Allowed suffices : None

Default suffix : None

:UP

Step up through stores. Use this command for memory sequencing

:DN

Step down through stores. Use this command for memory sequencing

Data type : None

Allowed suffices : None

Default suffix : None

Examples: RCL:FULL 125

RCL:UP

RCL?

Prepares message containing information on last memory store that was
recalled in the following format:

:RCL:MEM <nr1>

Examples: :RCL:MEM 126