YSI 9300 Photometer User Manual
Page 103
 
SEPARATION OF RESIDUALS 
The photometer is programmed for both zinc and the copper correction 
procedure. Use program Phot 35 Zinc (+ Copper), then select the ‘Follow On’ 
option on screen to continue test for program Phot 36 Corrected Zinc. The 
corrected zinc value is calculated automatically.
 
Test Procedure 
1 Fill test tube to the 10 ml mark. 
2 IN THE CASE OF CHLORINE CONTAINING SAMPLE ONLY :- 
Add one Zinc-Dechlor tablet, crush and mix to dissolve.
3 Add one Zinc tablet, crush and mix to dissolve. 
4 Allow the sample to stand for five minutes then mix again to ensure complete 
dissolution of the indicator.
5 Select Phot 35 on photometer. 
6 Take photometer reading in usual manner (see photometer instructions). 
The result is displayed as mg/l Zn.
7 FOR COPPER CONTAINING SAMPLES ONLY :-
Continue the test on the same test portion. Select the ‘Follow On’ option on
screen to continue the test program.
8 Add one EDTA tablet, crush and mix to dissolve. 
9 Take photometer reading in usual manner. 
10 The photometer displays the corrected zinc concentration as mg/l Zn.