Dissolved oxygen/2 – YSI 9300 Photometer User Manual

Page 125

Advertising
background image

PHOTOMETER TEST INSTRUCTIONS

_________________________________________________________________________________________

DISSOLVED OXYGEN/2

The presence of dissolved oxygen in water used in boilers and steam raising
plant is wholly undesirable. At high temperatures even small amounts of

dissolved oxygen render the water highly corrosive to boiler plant, the oxygen
must be removed from boiler feedwater by chemical treatment or by mechanical

de-aeration.
The DO/2.0 tests provide a simple means of testing for dissolved oxygen in
boiler water or boiler feedwater. The test covers the range and 0 - 2.0 mg/l.

Method

Special techniques must be employed when testing for dissolved oxygen as the

water sample can be readily contaminated by the oxygen in the atmosphere.
CHEMetrics Vacu-Vials self-filling reagent ampoules provide the ideal means of

carrying out this test. The tip of the ampoule is dipped into the water sample
and is then broken to allow the vial to fill with water. In this way there is no
possibility of contamination from the air.
The Vacu-Vial DO/2.0 test uses a reagent based on the indigo carmine method.
Indigo carmine, in its reduced form, reacts with dissolved oxygen to form a blue

complex.
The intensity of the colors formed in the tests is proportional to the dissolved
oxygen content of the water and is measured using a YSI Photometer.

Reagents and Equipment

DO Vial Set DO/2.0 (503)

YSI 9300 or 9500 Photometer

DO test reagents are light sensitive. Store tubes in original containers and keep

the box closed when not in use.

Photometer Method

AUTOMATIC
WAVELENGTH

SELECTION

0 – 2.0 mg/l

PHOT.50.AUTO

Advertising
This manual is related to the following products: