YSI 9300 Photometer User Manual

Page 40

Advertising
background image

Reagents and Equipment
YSI DPD No 1 Tablets
YSI DPD No 3 Tablets

YSI DPD Glycine Tablets
YSI DPD Acidifying Tablets

YSI DPD Neutralising Tablets
YSI 9300 or 9500 Photometer

YSI Round Test Tubes, 10 ml glass (PT 595)

Test Procedure - Chlorine Dioxide
1 Rinse a clean test tube with sample, then fill with sample to the 10 ml mark.

Add one Glycine tablet, crush and mix to dissolve.

2 Decant two or three drops of Glycine treated sample into a second clean test

tube. Add one DPD No 1 tablet and crush to disintegrate.

3 Add the remaining contents of the first test tube to the second test tube and mix.
4 Select Phot 7 on photometer.
5 Take photometer reading (Result G) immediately in usual manner (see

photometer instructions).

6 Multiply Result G by 5 to obtain the chlorine dioxide residual in terms of mg/l

Chlorine. To obtain the chlorine dioxide residual as mg/l ClO

2

, multiply

Result G by 1.9.


Test Procedure - Free and Combined Chlorine, and Chlorite
1 Rinse a test tube with sample leaving two or three drops. Add one DPD No 1

tablet, crush and then fill the test tube with sample to the 10 ml mark. Mix to
dissolve tablet.

2 Take the photometer reading on Phot 7 immediately (as result may drift on

standing), in usual manner (Result A).

3 Continue the test by adding one DPD No 3 tablet. Crush tablet, mix to dissolve

and then stand for two minutes.

4 Take photometer reading (Result C).
5 Continue the test by adding one DPD Acidifying tablet. Crush tablet, mix to

dissolve and then stand for two minutes.

6 Add one DPD Neutralising tablet, crush and mix to dissolve.
7 Take the photometer reading (Result D).
The results of the tests, in terms of mg/l chlorine, are calculated from the

observed results as follows:-

Chlorine Dioxide

= 5G

Free Chlorine

= A - G

Combined Chlorine

= C - A

Chlorite

= D - (C + 4G)

Total Oxidising Capacity = D