YSI 9300 Photometer User Manual

Page 60

Advertising
background image

Test Procedure
1 Filter sample if necessary to obtain a clear solution.
2 Fill the test tube with sample to the 10 ml mark.
3 Add one Calcicol No 1 tablet, crush and mix to dissolve.
4 Add one Calcicol No 2 tablet, crush and mix to dissolve.
5 Stand for two minutes to allow full color development.
6 Select Phot 12 on the photometer for result as mg/l CaCO

3

, or Phot 60 for result

as mg/l Ca.

7 Take photometer reading in the usual manner (see photometer instructions).

Interferences
1 Magnesium hardness (up to 200 mg/l as CaCO

3

) does not interfere with the

test.

2 Iron at levels above 10 mg/l may cause low results. Zinc above 5 mg/l may

cause high results.

3 The pH required in the test is closely controlled by a buffer mixture included in

the tablet formulation. However, to avoid exceeding the buffer capacity,
strongly acid or alkaline samples should be adjusted to within the pH range 4 to

10, prior to the start of the test.

Notes
1 The expression of hardness results sometimes causes confusion. It is normal

practice to express the results of hardness tests as mg/l CaCO

3

(calcium

carbonate). This is merely a convention to allow the comparison of different

results and does not necessarily indicate that the hardness is present in the
water in this form.

Results may also be expressed as mg/l Ca. To convert mg/l CaCO

3

to mg/l Ca

multiply by 0.4.

2 Magnesium hardness may be determined using the YSI Magnecol method (see

PHOT.21), or by taking the difference between the Total Hardness (PHOT.15)
and Calcium Hardness test results.

__________________________

Advertising
This manual is related to the following products: