YSI 9300 Photometer User Manual

Page 20

Advertising
background image

17

Reading in Transmittance and Absorbance
When taking readings in Transmittance or Absorbance mode, use the  and

keys to step through the wavelengths until the required wavelength is

reached.

Timer

The photometer features a countdown timer with alarm as an aid to carrying

out test procedures. The timer can be accessed at any time by selecting
‘Timer’ from the ‘Options’ menu.

The following display will appear:



Key in the time required in minutes and seconds using the numerical keys,
then press [OK] to start the timer. The maximum time is 29 minutes and

59 seconds. Use the  and  keys to reposition the cursor and re-enter the
time if it is keyed in incorrectly.

The timer will count down, giving an audible alarm at the end of the timed

period. Press [OK] to stop the alarm.

During the timer countdown period, an ‘Options’ menu is available :
Stop

- used to abort the timing operation or to stop the alarm at

the end of the timed period

Exit

- used to return to the program screen to take readings. The

timer will continue to run and give an audible alarm at the
end of the period.

Exit and Read - used to return to the program screen with the timer counting

down on screen. The instrument will automatically take a
reading at the end of the timed period and no alarm will

sound.

User Defined Tests (9500 Only)
Users may wish to develop their own test methods and store the calibration
data on the 9500. This will allow direct reading of user tests. The 9500 has
the facility to store up to 30 user-defined calibrations.

T im er

00:00

EXIT