YSI 9300 Photometer User Manual

Page 109

Advertising
background image

PHOTOMETER TEST INSTRUCTIONS

__________________________________________________________________________________________

Hydrazine is used as an oxygen scavenger in high pressure boilers and steam
raising plant. Hydrazine is particularly advantageous in that it does not

contribute solids to the boiler water.
The YSI Hydrazine test provides a simple means of measuring hydrazine levels

in boiler feed water and boiler water over the range 0 - 0.5 mg/l.

Method
The YSI Hydrazine test uses a special reagent powder containing p-dimethyl-
aminobenzaldehyde in an acidic formulation. Hydrazine reacts with this reagent to

produce a yellow coloration. The intensity of the color produced is proportional to
the hydrazine concentration and is measured using the YSI Photometer.

Reagents and Equipment

YSI Hydrazine Test Powder

YSI 9300 or 9500 Photometer
Round Test Tubes, 10 ml glass (PT 595)
Scoop, 1g approx (PT 697)

Test Instructions

1 Filter sample if necessary to obtain clear solution.
2 Take two test tubes A and B.
3 Fill test tube A with sample to the 10 ml mark.
4 Fill test tube B with deionised water to the 10 ml mark.
5 To each tube add one level scoop (1g) of Hydrazine Test Powder. Mix to

dissolve and cap tubes.

6 Stand for two minutes to allow full color development.
7 Select Phot 41 on photometer.
8 Take photometer reading of Tube A in usual manner (see photometer instructions).
9 Use Tube B as the Blank to set the instrument.

10 The result is displayed as mg/l N

2

H

4

.

Photometer Method
AUTOMATIC
WAVLELENGTH

SELECTION
0 – 0.5 mg/l N

2

H

4

PHOT.41.AUTO

Advertising
This manual is related to the following products: