Features and Technical Specification

Application

For application in general water testing using Palintest
tablet reagent systems and Palintest Tubetests reagents.

Instrument Type

Single-beam colorimeter with built-in colour filters and pre-
programmed test calibrations.

Peak Wavelengths

445

5nm, 495

5nm, 555

5nm, 570

5nm, 605

5nm and 655

5nm

Range

1 - 100%T

Accuracy

1%T

Display

Large graphic display with option of backlight.

Language

Test identification and prompts in English, French,

German, Spanish and Italian.

Timer

Clock and timer feature to log test results and audible

alarm for timing test procedure.

Units

Direct-reading of test results in mg/l, ppm, g/l or molar
units (mmol/l or µmol/l).

User Selectable Options

10 digit sample number entry, dilution factor, time/date,
date format, system lock and rounding of results.

Date Format

Date format selectable as day/month/year or month/day/
year.

Zeroing

Automatic zeroing on blank tube and hold blank facility for
series of tests. Continuation test facility without the need

for reblanking.

Internal Memory

Stores 500 previous readings with option to view logged

results on screen, or download to computer.

USB Interface

USB 1.1 full-speed, bus-powered device. Software selectable

between either emulation of a removable hard-drive or
emulation of a serial device connected via a virtual COM port.

Power

3 x 1.5V ‘AA’ alkaline batteries or via USB interface. Power
management system with variable length auto switch-off

or 'continuous' operation.

Size

250 x 150 x 70 mm

Weight

985 g (2.1 lbs)

Test Tubes

For tablet reagents - 10 ml glass test tubes, 20 mm OD
(YPT 595).

Cell Holder

Multi-size tube holder accepts test tubes from 12 –
20 mm OD and centres the tube for optimum optical
performance.