YSI 9300 Photometer User Manual

Page 78

Advertising
background image

TEST PROCEDURE
1 Using the measuring syringe take a 1 ml sample of the water under test.

Transfer to the round test tube and make up to the 10 ml mark with
deionised water.

2 Add one Magnecol tablet, crush and mix to dissolve.
3 Stand for five minutes to allow full color development and the slight turbidity

to clear.

4 Select Phot 21 on photometer for result as mg/l Mg. Select Phot 61 for result

as magnesium hardness, mg/l CaCO

3

.

5 Take photometer reading in usual manner (see photometer instructions).

Note
: To convert mg/l Mg to magnesium hardness as CaCO

3

, multiply by 4.2.

_________________________