Sample Collection
To prevent loss of sulphide collect the sample carefully with a minimum of
agitation or aeration. Test the sample as soon as possible after collection.

Test Procedure

1 Fill test tube with sample to the 10 ml mark.
2 Add one Sulphide No 1 tablet and one Sulfide No 2 tablet. Crush and mix

gently to dissolve the tablets. Gentle mixing is essential to avoid loss of
sulphide.

3 Stand for 10 minutes to allow full color development.
4 Select Phot 33 on photometer.
5 Take photometer reading in usual manner (see photometer instructions).
6 The result is displayed as mg/l S.

To convert from mg/l S to mg/l H

2

S - multiply result by 1.06