YSI 9300 Photometer User Manual

Page 130

Advertising
background image

Test Procedure

1

Fill test tube with sample to the 10 ml mark.

2

Add one Nickeltest No 1 tablet, crush and mix to dissolve. Ensure tablet is

completely dissolved before proceeding.

3

If iron is thought to be present in the sample, add one level spoonful of

Nickeltest PR powder and mix.

4

Add one Nickeltest No 2 tablet, crush and mix to dissolve.

5

Stand for two minutes to allow full color development.

6

Select Phot 53 on photometer.

7

Take photometer reading in usual manner (see photometer instructions).

8

The result is displayed as mg/l Ni.

Interferences

1

The presence of cobalt at 0.5 mg/l gives a positive response in the test.

2

The presence of significant levels of EDTA (at least 25 mg/l) complexes

nickel and reduces response in the test. Complexing agents used in water

treatment, such as polyphosphates, do not affect the results.