Test Procedure

1 Fill test tube with sample to the 10 ml mark.
2 Add one Cyanuric Acid tablet and allow to disintegrate for at least two

minutes. A cloudy solution indicates the presence of cyanuric acid.

3 Crush any remaining undissolved tablet and mix to ensure uniformity.
4 Select Phot 13 on photometer.
5 Take photometer reading in usual manner (see photometer instructions).
6 The result is displayed as mg/l.

Note

The range of the test is 0 - 200 mg/l. However when a test result of 100 mg/l or

over is obtained, the following dilution technique is recommended in order to
obtain a more precise result.
1 Take a sample of pool water in a YSI Dilution Tube (PT 512), filling to the

x10 mark.

2 Make up to the 'Deionised Water' mark with deionised water, or tap water,

and mix.

3 Fill a round test tube with diluted sample to the 10 ml mark. Test as per the

earlier test procedure.

4 Multiply the displayed result by 10 to obtain the cyanuric acid concentration.