YSI 9300 Photometer User Manual

Page 97

Advertising
background image

Reagents and Equipment
YSI Sulfate Turb Tablets
YSI 9300 or 9500 Photometer
Round Test Tubes, 10 ml glass (PT 595)

Test Procedure
1 Fill test tube with sample to the 10 ml mark.
2 Add one Sulfate Turb tablet, crush and mix to dissolve. A cloudy solution indicates

the presence of sulphate.

3 Stand for five minutes then mix again to ensure uniformity.
4 Select Phot 32 on photometer.
5 Take photometer reading in usual manner (see photometer instructions).
6 The result is displayed as mg/l SO

4

.

Caution
YSI Sulfate (Turb) tablets each contain 20 mg Barium Chloride. These tablets are
harmful if ingested. Avoid handling tablets whenever possible and wash hands

after use.

_________________________

Advertising