Test Procedure 
Read the Oxygen test instructions leaflet contained in the CHEMetrics Vacu-Vials 
pack. Observe these various recommendations regarding sample handling and 
use of Vacu-Vials :- 
1 Carry out the test in accordance with the test procedure given in the 
CHEMetrics instruction leaflet. Observe the time periods given in the test
instructions.
2
Select Phot 49 on the photometer. The wavelength is set automatically.
3
Use the colorless blank ampoule provided in the Starter Pack as the blank for
setting the instrument.
4
The photometer provides direct readings in mg/l O
2
.
 
Note 
Note for measuring dissolved oxygen in natural waters, use the DO 15/20 test 
(see instructions Phot.75). 
Vacu-Vials is a registered trade mark of CHEMetrics Inc