Test Procedure - Total Chromium

1

Fill the pretreatment tube with sample to the 20 ml mark.

2

Add one level spoonful of Chromicol CR powder. Replace screw cap and

shake tube well for two minutes.

3

Allow tube to stand for two minutes.

4

Filter a portion of the solution dropwise into a round glass test tube. Discard

the first few drops and then fill to the 10 ml mark.

5

Add one Chromicol No 1 tablet, crush and mix to dissolve.

6

Add one Chromicol No 2 tablet, crush and mix to dissolve.

7

Stand for 10 minutes without disturbing the solution to allow full color

development and to enable any undissolved particles to settle.

8

Select Phot 100 on photometer.

9

Take photometer reading in usual manner (see photometer instructions).

10

The result represents the total soluble chromium concentration (trivalent and

hexavalent) as mg/l Cr (Result B).

11

The trivalent Chromium (Cr

III

) concentration is obtained by subtracting Result

A from Result B :-

Trivalent Chromium = Result B - Result A


Interferences

Levels of dissolved iron above 1 mg/l cause low results for chromium. To increase

the tolerance, repeat the test using two Chromicol No 1 tablets and one
Chromicol No 2 tablet. Tannin causes complexation which prevents a response in

the test.