YSI 9300 Photometer User Manual

Page 95

Advertising
background image

Test Instructions
1 Fill test tube with sample to the 10 ml mark
2 Add one Silica No 1 tablet, crush and mix to dissolve. Stand for five minutes to

allow the silica to react.

3 Add one Silica PR tablet, crush and mix to dissolve. (This stage may be

omitted if the sample is known to be completely free of phosphate).

4 Add one Silica No 2 tablet, crush and mix to dissolve. Stand for one minute

to allow full color development.

5 Select Phot 31 on photometer.
6 Take photometer reading in usual manner (see photometer instructions).
7 The result is displayed as mg/l SiO

2

.

Note
For testing high levels of Silica the YSI Silica HR test should be used. The range
of this test is 0 - 100 mg/l (see Test Instruction Phot 56).