YSI 9300 Photometer User Manual

Page 132

Advertising
background image

Reagents and Equipment
YSI Phenoltest No 1 Tablets
YSI Phenoltest No 2 Tablets
YSI Phenoltest PR Tablets

YSI 9300 or 9500 Photometer
Round Test Tubes, 10 ml glass (PT 595)

Test Procedure

1

Fill round test tube to the 10 ml mark with sample.

2

In the case of samples known to contain copper, zinc, iron or manganese
ions, add one Phenoltest PR tablet. Crush and mix to dissolve.

3

Add one Phenoltest No 1 tablet, crush and mix to dissolve.

4

Add one Phenoltest No 2 tablet, crush and mix to dissolve.

5

Stand for 10 minutes to allow full color development.

6

Select Phot 54 on the photometer.

7

Take photometer reading in the usual manner (see photometer instructions).

8

The result is displayed as mg/l C

6

H

5

OH.


Interferences

1

Use of the Phenoltest PR tablet will prevent interference from metal ions up

to a concentration of 350 mg/l. The test is unaffected by free chlorine in the
sample up to 10 mg/l.

2

Low results may be obtained in samples containing more than 150 mg/l

alkalinity (as CaCO

3

), 10 mg/l sulphite or 2 mg/l sulphide. Certain organic

keto-enol compounds may cause high results. In the case of known or
suspected interferences, then the sample should be pre-treated in accordance

with standard analytical procedures.