Reagents and Equipment

YSI Acidifying CD Tablets
YSI Chloridol Tablets
YSI 9300 or 9500 Photometer

Round Test Tubes, 10 ml glass (PT 595)
Measuring Syringe, 1 ml (PT 361)

Sample Container, 100/50/10 ml plastic (PT 510)


Test Calibration
Select Program Phot 46

Range 0 – 50 mg/l Cl

or Phot 51

Range 0 – 500 mg/l Cl

or Phot 101

Range 0 – 10,000 mg/l NaCl

or Phot 102

Range 0 – 50,000 mg/l NaCl

Test Instructions
For Testing Boiler Condensate and Softened Waters ONLY

Range 0 - 50 mg/l Cl

1 Fill test tube with sample to the 10 ml mark.

For Testing Natural Waters, Drinking Water, Swimming Pools and Boiler
Waters

Range: 0 - 500 mg/l Cl

Using the measuring syringe, take 1 ml of sample. Transfer to the test tube

and make-up to the 10 ml mark with deionised water.

For Testing Salt Chlorinator Treated Swimming Pools

Range: 0 - 10,000 mg/l NaCl

Using the measuring syringe, take 0.5 ml of sample. Transfer to the sample
container (PT 510) then make-up to the 100 ml mark with deionised water.

Cap tube and mix. Fill test tube to the 10 ml mark with solution from the
sample container.

For Testing Sea Water and Brackish Waters

Range: 0 - 50,000 mg/l NaCl

2 Using the measuring syringe, take 0.1 ml of sample. Transfer to the sample

container (PT 510) then make-up to the 100 ml mark with deionised water.

Cap tube and mix. Fill test tube to the 10 ml mark with solution from the
sample container.

3 Add one Acidifying CD tablet, crush and mix to dissolve.

4 Add one Chloridol tablet, allow the tablet to disintegrate for two minutes then

crush any remaining particles and mix. A cloudy solution indicates the presence
of chloride.