Test Procedure

1

Filter sample if necessary to obtain a clear solution.

2 Using the measuring syringe take 1 ml of the sample. Transfer to the test tube

and make up to the 10 ml mark with deionised water.

3

Add one Nitriphot No 1 tablet, crush and mix to dissolve.

4

Add one Nitriphot No 2 tablet, crush and mix to dissolve. Cap immediately.

5

Stand for exactly two minutes to allow full color development. Ignore any

further color development after this time.

6

Select Phot 43 on the photometer.

7

Take photometer reading in the usual manner (see photometer instructions).

8

The result is displayed as mg/l NaNO

2

.

Interferences

Chlorine in excess of 30 mg/l may give slight positive interference. However,
nitrite and chlorine are incompatible and do not normally co-exist.
The solution should be cooled to below 30°C before testing for the most accurate
analytical results.