YSI 9300 Photometer User Manual

Page 80

Advertising
background image

Reagents and Equipment
YSI Molybdate No 1 HR Tablets
YSI Molybdate No 2 HR Tablets
YSI 9300 or 9500 Photometer

Round Test Tubes, 10 ml glass (PT 595)

Test Procedure

1 Fill round test tube with sample to the 10 ml mark.
2 Add one Molybdate No 1 HR tablet, crush and mix to dissolve.
3 Add one Molybdate No 2 HR tablet, crush and mix to dissolve.
4 Select Phot 22 on photometer.
5 Take photometer reading in usual manner (see photometer instructions).

6 The result is displayed as mg/l MoO

4

.

Note
Molybdate concentrations can be expressed in a number of different ways. The
following factors may be used for the conversion of readings :-

To convert from MoO

4

to Na

2

MoO

4

- multiply by 1.3

To convert from MoO

4

to Mo - multiply by 0.6

Interferences
The presence of Fe

2+

(ferrous) or Fe

3+

(ferric) iron at 5 mg/l causes the rapid

development of a red color in the test. This may be prevented by the treatment

of the 10 ml sample with one YSI EDTA tablet prior to carrying out the test.