MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-57

6. Open the transducer’s stopcock to air to zero the transducer and remove stopcock cap. Select

the

P1

area. Select

ZERO

from the menu. The message

P1 ZEROED

appears when zeroing is

complete and the pressure values are displayed as zeros.

7. Close the stopcock to air. The patient’s pressure waveform should be displayed. A scale is

automatically selected to display the pressure. Confirm that pressure amplitude correlates with

the digital readout.

Note:

If you place a cap on an open port before you close the port to air, an error message may

appear. You will be required to zero the transducer again.

If pressure alarms are desired, set the alarms after you obtain a satisfactory waveform. Error or

alarm messages appear in the message area at the bottom of the screen. For more information, see

"Alarms" on page 3-21.

IP Scale Options

The IP monitor can display pressures from -30 to 300 mmHg. After zeroing the transducer

pressure, the monitor automatically selects one of the following scales based on the patient’s

measured pressure:

• -30 to 30 mmHg

• 0 to 60 mmHg

• 0 to 120 mmHg

• 0 to 150 mmHg

• 0 to 180 mmHg

• 0 to 300 mmHg

You can also manually select one of these scales or autoscale to readjust the waveform within the

channel.

To change the scale:

1. Use the

SPEED DIAL

to outline and select the P1 area. The P1 menu appears.

2. From the menu, select

SCALE

and then choose a scale from the list.