Lifepak, 15 monitor/defibrillator operator’s checklist – Physio-Control LIFEPAK 15 User Manual

Page 287

Advertising
background image

LIFEPAK

®

15 Monitor/Defibrillator Operator’s Checklist

This is a recommended checklist to use to inspect and test this monitor/defibrillator. Daily inspection
and test is recommended. This form may be reproduced.

Unit Serial No:______________________________ Location:_____________________________________

Instruction

Recommended

Corrective Action

Date

Initials

✔ each box after completing

1. Inspect physical condition for:

Foreign substances

Clean the device.

Damage or cracks

Contact a qualified service technician.

2. Inspect power source for:

Broken, loose, or worn battery pins

Contact a qualified service technician.

Damaged or leaking battery

Recycle or discard battery.

Spare battery available

Obtain fully charged spare battery.

Damage to power adapter and cables

Contact a qualified service technician.

3. Inspect ECG cable and cable port for:

Cracking, damage, broken, or bent
parts or pins

Replace ECG cable.
If port is damaged, contact qualified
service technician.

4. Check ECG electrodes and therapy electrodes for:

Use By date

Replace if date passed.

Spare electrodes available

Obtain spare electrodes.

Damaged, opened package

Discard and replace electrodes.

5. With batteries installed, disconnect from power adapter (if using), press ON and observe for:

Momentary illumination of
self-test messages and LEDs, and
speaker beep

If absent, contact a qualified
service technician.

Two fully charged batteries

Replace low battery or charge installed
battery using power adapter.

Service indicator (

)

If illuminated, contact a qualified service
technician.

6. With batteries installed, reconnect power adapter to device and check for:

(If not using a power adapter, go to Step 7.)

Power adapter LED strip is illuminated If absent, check connections.

If problem persists, contact a
qualified service technician.

Auxiliary power LED on device is
illuminated

Battery charging LED on device is
illuminated or flashing

If absent, check batteries. If problem
persists, contact a qualified service
technician.

7. Perform QUIK-COMBO

®

therapy cable check in Manual mode:*

(If this cable is not used with the defibrillator, go to Step 8.)

a.Disconnect and examine cable for

cracking, damage, broken, or bent
parts or pins.

Replace QUIK-COMBO therapy cable.

b.Connect therapy cable to defibrillator

and the Test Load.

If

CONNECT ELECTRODES

,

PADDLES LEADS OFF

,

CONNECT CABLE

, or

ABNORMAL ENERGY

DELIVERY

message appears,

replace therapy cable and repeat check.
If problem continues, remove the
defibrillator from use and contact a
qualified service technician.

c.Select

LEAD

, then

PADDLES

.

d.Select

200 JOULES

and press

CHARGE

.

e.Press (shock)

button.

f. Confirm

ENERGY DELIVERED

message appears.

If message does not appear, replace
therapy cable and repeat check.

g.Remove Test Load from cable and

verify

PADDLES LEADS OFF

appears.**

If absent, contact a qualified service
technician.