LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Troubleshooting Tips

LOST DIRECT CONNECTION

message appears

Direct connection was interrupted

• Verify cable connections between

LIFEPAK 15 monitor and gateway
or PC.

• Attempt to retransmit.

LOST BLUETOOTH
CONNECTION

message

appears

Connection with Bluetooth target
device was interrupted

• Verify target device is ready to

receive transmissions.

• Attempt to retransmit.

TRANSMISSION CANCELLED

message appears

Operator of the LIFEPAK 15
monitor cancelled transmission

• Attempt to retransmit if cancelled

in error.

Table 8-3 Troubleshooting Tips for Data Transmission (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION