Automated external defibrillation (aed) – Physio-Control LIFEPAK 15 User Manual

Page 134

Advertising
background image

5-20

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Automated External Defibrillation (AED)

DISARMING

message appears

(energy charge removed)

(shock)

button not pressed

within 60 seconds after charge
complete

• Recharge the defibrillator, if

desired.

SPEED DIAL

pressed

• Recharge the defibrillator.

Therapy electrodes or cable
disconnected

• Reconnect electrode or cable.

Energy did not escalate

After a shock, the next
analysis was NO SHOCK ADVISED

• No action needed.

Defibrillator does not
escalate energy when a NO
SHOCK ADVISED

decision

follows a shock.

Charge time to 360 joules
exceeds 10 seconds

Battery low

• Replace battery with fully

charged battery.

• Connect to auxiliary power

using approved power
adapter.

Operating temperature is too
low

• Move patient and device to

warmer environment, if
necessary.

REPLACE BATTERY

prompt occurs

Both batteries are very low

• Replace one or both batteries

immediately.

• Connect to auxiliary power

using approved power
adapter.

Voice prompts sound faint or
distorted

Low battery power

• Replace one or both batteries

immediately.

• Connect to auxiliary power

using approved power
adapter.

CPR time shown (minutes/
seconds) is different than
expected

Function of metronome

• None. The metronome

adjusts the CPR time to
ensure CPR cycle ends with
compressions. (See
page 5-14.)

Incorrect setup option
selected

• Change CPR time setup

option. See LIFEPAK 15
Monitor/Defibrillator Setup
Options
provided with your
device.

Press CPR and metronome does
not activate

In AED mode, and not in CPR
interval

• Wait until CPR interval

(audible “tocks”) to silence
or activate metronome.

Table 5-1 Troubleshooting Tips for AED Mode (Continued)

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION