3-20

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Home Screen

Battery Status Indicators

The Home Screen displays battery indicators that show the following information about the

batteries installed in the defibrillator:

• Presence or absence of battery in battery well

• Battery in use

• Battery charge state

IMPORTANT!

Always check the battery charge level and ensure batteries are adequately charged

before use.

When two batteries are installed, the defibrillator uses the battery with the lowest level of charge

first. The battery in use is indicated by a white battery number in a black box. When a battery

reaches the replace battery state, the defibrillator automatically switches to the other battery. When

all battery capacity is exhausted, the defibrillator turns off. If you insert a charged battery and

repower the device in less than 30 seconds, the defibrillator retains its settings. Table 3-8 provides

a description of the various battery status indicators.

Note:

When the defibrillator is operating on auxiliary power using a power adapter, the battery

indicators show the battery charge level, but the well numbers are not highlighted. The

LOW BATTERY

and

REPLACE BATTERY

messages and prompts do not occur when operating on auxiliary power.

Table 3-8 Battery Status Indicators

INDICATOR

MEANING

DESCRIPTION

Active battery

The defibrillator is using the battery in well 1 for power. Battery status
indicators display up to four green bars. Each green bar represents
approximately 25% remaining charge. For example, three green bars
indicate about 75% remaining charge.

Low battery

Battery in well 1 is in use and is low. One yellow bar indicates 5% to
10% remaining charge.

Very low
battery

Battery in well 1 is in use and is very low. One red flashing bar
indicates 0 to 5% remaining charge. The defibrillator automatically
switches to the other battery only if adequate charge is available. If
both batteries show red bars, the REPLACE BATTERY voice prompt
occurs.

Unrecognized
battery

Battery in well 2 is not in use. Battery communication failed or a non-
Physio-Control battery is installed. The battery may power the
defibrillator but the level of charge is unknown and low battery
messages and prompts will not occur.

No battery
installed or
fault
detected

No battery is installed in battery well 1, or a fault was detected in the
battery in well 1 and the device will not use the battery.