LIFEPAK 15 Monitor/Defibrillator Operating Instructions

5-13

Subsequent analysis for

SHOCK ADVISED

and

NO SHOCK ADVISED

sequences are the same as

described above. The energy level for Shock 2, 3, and greater depends on the

ENERGY PROTOCOL

setup and the analysis decision. When a

NO SHOCK ADVISED

decision follows a shock, the energy

level does not increase for the next shock. When a

SHOCK ADVISED

decision follows a shock, the

energy level increases for the next shock.

Motion Detected

Electrodes or Therapy Cable Off

AED Mode

If the AED detects motion during the ECG
analysis, the MOTION DETECTED, STOP MOTION
prompts occur, followed by a warning tone.

Analysis is inhibited until the motion stops
or for up to 10 seconds. After the motion
ceases or 10 seconds have elapsed, analysis
continues to completion even if motion is
still present. For possible causes of motion
detection and suggested solutions, see
Table 5-1 on page 5-19.

Motion detected!

Stop motion!

AED Mode

Connect electrodes

If therapy electrodes are not connected, the
CONNECT ELECTRODES

prompts occur until the

patient is connected.

AED Mode

Connect cable

If the therapy cable is not connected to the
defibrillator, the CONNECT CABLE message
appears until the cable is connected.