Displaying and printing trend graphs, Vs and st monitoring considerations, Vital sign and st segment trends – Physio-Control LIFEPAK 15 User Manual

Page 114

Advertising
background image

4-70

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Vital Sign and ST Segment Trends

Displaying and Printing Trend Graphs

The trend graph for any active vital sign or ST measurement can be displayed in Channel 2 or 3.

The example in Figure 4-18 shows the trend graph in Channel 3. Only two trend graphs can be

displayed at a time, but the device collects trend data on all active vital sign values.

To display trend graphs:

1. Rotate the

SPEED DIAL

to outline Channel 2 or 3, and then press the

SPEED DIAL

to select the

channel. The Channel menu appears.

2. Select

WAVEFORM

, and then select

TREND

.

3. Select

SOURCE

, and then select the desired VS or ST.

4. The default setting for

SCALE

and

RANGE

is

AUTO

. When

AUTO

is used, the monitor automatically

updates the scale so that all values are displayed and all data from Power On to the present time

is visible. If you change scale or range, some data may not be visible because it is off scale or

out of range.

5. Press

HOME SCREEN

. The graph for the selected VS or ST appears in the channel.

Note: To initiate ST trends, you must obtain a 12-lead ECG. The initial ECG provides the baseline

ST measurement and initiates the ST trends feature.

To print trend graphs:

1. Press

OPTIONS

. The Options menu appears.

2. Rotate and then press the

SPEED DIAL

to select

PRINT

.

3. Select

REPORT

, and then select

TREND SUMMARY

.

4. Select

PRINT

. The Trend Summary Report prints graphs of all actively monitored VS and ST

trends.

VS and ST Monitoring Considerations

For best results, consider the following:

• The ability of the patient to cooperate and be relaxed. Patients who are restless can produce

noisy physiological signals. Noisy signals can result in inaccurately high or low data

measurements.

• The quality of the physiological signal. If the ECG has significant artifact, the HR may have

spurious measurements. Noisy 12-lead ECGs may need to be overridden, and ST measurements

will not be obtained.

• The expected length of time the patient is to be monitored. VS graphs of the patient monitored

for only a short time (for example, 15 minutes) may not provide enough data to identify gradual

changes in patient condition.

• The patient ECG rhythm. Diagnosis of ST associated ischemia is inhibited by certain ECG

findings such as left bundle branch block and ventricular pacing.

Advertising