Initial cpr - analyze first, Automated external defibrillation (aed) – Physio-Control LIFEPAK 15 User Manual

Page 130

Advertising
background image

5-16

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Automated External Defibrillation (AED)

• If you did witness the arrest, press

ANALYZE

. The CPR period ends, and the

ANALYZING NOW,

STAND CLEAR

prompts occur.

• If you did not witness the arrest, perform CPR and do not press

ANALYZE

. The Initial CPR

countdown timer continues for the duration specified in the

INITIAL CPR TIME

setup option, for

example, 90 seconds. When initial CPR time ends, the

PUSH ANALYZE

prompts occur.

Initial CPR - Analyze First

When the

INITIAL CPR

option is set to

ANALYZE FIRST

, you are prompted to perform analysis after

the AED is turned on. CPR is prompted after the AED completes the analysis.

If the electrodes are not attached to the patient, the

CONNECT ELECTRODES

prompts occur before

you are prompted to perform analysis.

No Shock Advised

If the AED detects a nonshockable rhythm, the

START CPR

prompts occur.

AED Mode

Start CPR

1:46

The START CPR prompts occur.

AED Mode

Start CPR

1:46

If you witnessed the arrest, push ANALYZE

After 3 seconds, a countdown timer
appears and the IF YOU WITNESSED THE
ARREST, PUSH ANALYZE

prompts occur.

These prompts provide an opportunity to
end the initial CPR early and proceed
directly to analysis.

Note: The decision to end CPR early is
based on your protocol and if you witnessed
the arrest.

AED Mode

Start CPR

1:46

A countdown timer (min:sec format)
continues for the duration specified in the

INITIAL CPR TIME

setup option.