Lead ecg procedure, Monitoring – Physio-Control LIFEPAK 15 User Manual

Page 61

Advertising
background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-17

To locate the V1/C1 position:

1. Place your finger at the notch in the top of the sternum.

2. Move your finger slowly downward about 3.8 centimeters (1.5 inches) until you feel a slight

horizontal ridge or elevation. This is the Angle of Louis where the manubrium joins the body

of the sternum.

3. Locate the second intercostal space on the patient’s right side, lateral to and just below the

Angle of Louis.

4. Move your finger down two more intercostal spaces to the fourth intercostal space, which is

the V1/C1 position.

5. Continue locating other positions from V1/C1 (see Figure 4-5).

Other important considerations:

• When placing electrodes on female or obese patients, always place leads V3-V6 and C3-C6

under the breast rather than on the breast.

• Never use the nipples as reference points for locating the electrodes for men or women

patients, because nipple locations vary widely.

12-Lead ECG Procedure

To acquire a 12-lead ECG:

1. Press

ON

.

2. Insert the lead attachments into the main cable as shown in Figure 4-6.

Figure 4-6 12-Lead ECG Cables

3. Insert the cable connector into the green ECG connector on the monitor.

4. Prepare patient’s skin for electrode application (see page 4-9).

5. Apply ECG electrodes (see page 4-16).

Protective flap

(open)

Main cable

(4-wire)

Precordial lead

attachment

Precordial lead

attachment

Limb lead

attachment

Main cable