Basic orientation, Power adapter – Physio-Control LIFEPAK 15 User Manual

Page 199

Advertising
background image

POWER ADAPTER

9

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

9-3

Basic Orientation

The AC Power Adapter and DC Power Adapter are optional accessories for use only with the

LIFEPAK 15 monitor/defibrillator. These power adapters:

• Provide operating power to the monitor/defibrillator with or without batteries installed.

• Provide power to charge batteries installed in the monitor/defibrillator.

The AC Power Adapter operates with either 120 or 240 Vac line power. The DC Power Adapter

operates with 12 Vdc power. Installed batteries are charged whenever the power adapter is

connected to the LIFEPAK 15 monitor/defibrillator. To help manage and maintain battery charge,

the power adapter should be kept plugged into the defibrillator whenever possible. For more

information about maintaining the batteries, see "Battery Maintenance" on page 10-12.

Note:

Although the monitor/defibrillator can operate using auxiliary power with no batteries

installed, at least one battery should be installed at all times.

Note:

If the monitor/defibrillator loses power for more than 30 seconds, it will revert to the user-

configured default settings and begin a new patient record.

An optional output extension cable is available. The output extension cable is equipped with a

breakaway connector to allow quick movement if needed. For more information about the

breakaway feature, see "Output Extension Cable with Breakaway Connector" on page 9-7.

IMPORTANT!

Daily inspection and testing will help ensure that the power adapter is in good

operating condition and is ready for use when needed. Refer to the LIFEPAK 15 monitor/

defibrillator Operator’s Checklist in the back of this manual.

Carefully read the Power Adapter Instructions for Use that are provided with the power adapter for

complete instructions, warnings, cautions, and specifications.