Troubleshooting tips, Power adapter – Physio-Control LIFEPAK 15 User Manual
Page 205

POWER ADAPTER
9
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
9-9
Troubleshooting Tips
Table 9-3 Troubleshooting Tips for Power Adapter
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
POWER
LED on power adapter
does not light
Power cord not plugged into
power adapter or power source
• Connect power cord.
Defective power adapter or
power cord
• Replace with working power
adapter and power cord.
Blown fuse or tripped circuit
breaker in building
• Contact qualified service
personnel.
AUXILIARY POWER
LED on
monitor/defibrillator not
illuminated
Power adapter not properly
connected to auxiliary power
source or monitor/defibrillator
• Check that power adapter is
connected properly.
Defective power adapter or
cables
• Replace with working power
adapter and cables.
BATTERY CHARGING
LED on
monitor/defibrillator not
illuminated
Power adapter not properly
connected to auxiliary power
source or monitor/defibrillator
• Check that power adapter is
connected properly.
Battery not properly inserted in
battery well
• Check that battery is properly
inserted in battery well.
Unable to charge battery with
power adapter because battery
charge level is too low
• Charge battery in Station-Mobile
or REDI-CHARGE battery charger
if available.
• Replace battery.
No batteries installed
• Install at least one battery.
Defective battery
• Remove battery from service and
replace with working battery.
Unrecognized battery
• Only use battery that is approved
for use with the LIFEPAK 15
monitor/defibrillator.
Incompatible power adapter
connected to the monitor/
defibrillator
• Only use power adapter that is
approved for use with the
LIFEPAK 15 monitor/defibrillator.
Defective power adapter or
cables
• Replace with working power
adapter and cables.
Monitor/defibrillator unable to
recognize installed battery
• Contact qualified service
personnel.