5-10

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Automated External Defibrillation (AED)

6. Press

ANALYZE

to initiate the analysis. Stop CPR.

7. Continue to follow the screen messages and voice prompts provided by the AED.

Shock Advised

The following prompts occur when shock is advised:

WARNING

POSSIBLE MISINTERPRETATION OF DATA

Do not move the AED during analysis. Moving the AED during analysis may affect the

ECG signal resulting in an inappropriate

SHOCK

or

NO SHOCK ADVISED

decision. Do not

touch the patient or the AED during analysis.

The ANALYZING NOWSTAND CLEAR prompts
occur. The SAS analyzes the patient’s ECG in
approximately 6 to 9 seconds and advises
either SHOCK ADVISED

or NO SHOCK ADVISED.

AED Mode

Analyzing now -- Stand clear

If the AED detects a shockable rhythm, the
SHOCK ADVISED

prompts occur. Charging to

the joule setting for Shock #1 begins. A
charging bar appears and a ramping tone
sounds.

AED Mode

SHOCK ADVISED!
Charging to 200J

200J

To cancel, push Speed Dial