How st trends work, Interpreting the st trend graph, Vital sign and st segment trends – Physio-Control LIFEPAK 15 User Manual

Page 112

Advertising
background image

4-68

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Vital Sign and ST Segment Trends

How ST Trends Work

ST measurements can be displayed graphically for time ranges of 30 minutes, and 1, 2, 4, and 8

hours. ST trending is initiated by obtaining the patient’s first 12-lead ECG. The ST J-point (STJ) is

the part of the ST segment that is measured (see Figure 4-17). The STJ measurement is plotted on

the ST trend graph (see Figure 4-18).

Figure 4-17 STJ Measurement

When all leads of the 12-lead ECG cable are attached to the patient, STJ measurements are

obtained automatically every 30 seconds. If a lead is off, or the ECG data is too noisy, ST

measurements are not obtained and the graph shows a blank for that time period. If an STJ

measurement in any lead deviates from the initial measurement by 1 mm (0.1 mV) or more and the

deviation persists for 2.5 minutes, the monitor automatically prints another 12-lead ECG. Manual

requests for 12-lead ECGs do not affect ST trending or automatic printing.

Interpreting the ST Trend Graph

Using the first 12-lead ECG, the monitor identifies the presence of any STJ displacement, either

negative or positive, and the lead that has the most STJ displacement. When

AUTO

is selected, the

lead that has the most STJ displacement is shown on the graph. The STJ is measured every 30

seconds thereafter.

Figure 4-18 shows an example of an ST trend graph. The elapsed time goes from right to left across

the screen. The most current STJ measurement is on the far right. Each time an STJ measurement

is obtained, it is compared to the first STJ or baseline measurement. The bars represent the change

in the STJ compared to the first measurement.

STJ

Advertising