Cleaning, Troubleshooting tips, Monitoring invasive pressure – Physio-Control LIFEPAK 15 User Manual

Page 102

Advertising
background image

4-58

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Monitoring Invasive Pressure

Cleaning

IP transducers are disposable and are intended for single-patient use. Do not clean and reuse

transducers. Dispose of the contaminated waste according to local protocols.

IP cables are reusable and may be cleaned. To clean the reusable IP cable:

1. Disconnect the cable from the monitor.

2. Use a clean, soft cloth dampened with a germicidal solution to wipe clean.

3. Allow to dry before reconnecting the cable to the monitor.

For information about cleaning the device, see "Cleaning the Device" on page 10-15.

Troubleshooting Tips

The error messages in Table 4-8 use the text

PX

to represent any of the labels for invasive pressure,

including P1, P2, and the user-selectable labels ART, PA, CVP, ICP, and LAP.

Table 4-8 Troubleshooting Tips for IP Monitoring

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

Invasive pressure value is blank

No transducer is connected

• Connect the transducer to the

cable, and the cable to the
monitor.

No scale appears next to the
waveform

The zero reference has not
been established

• Zero the transducer.

PX NOT ZEROED

message appears

The zero reference has not
been established

• Zero the transducer.

PX ZERO FAILED

message appears

An unsuccessful attempt
has been made to set a zero
reference value

• Make sure that the transducer

is open to air and repeat the
attempt to zero.

Dampened waveform

Loose connection

• Check the entire system for

leaks. Tighten all connections.
Replace any defective
stopcocks.

Tubing too long or too
compliant

• Use short, stiff tubing with

large diameter.

Thrombus formation, air
bubbles, or blood left in
catheter after blood draw

• Use syringe to draw back air or

particles in catheter, and then
flush system.

Kinked catheter, catheter
tip against vessel wall,
arterial spasm

• Reposition catheter. Anchor

catheter to skin at insertion
site.

Resonating waveform

Tubing too long

• Use short, stiff tubing with a

large diameter.