Troubleshooting tips, Monitoring – Physio-Control LIFEPAK 15 User Manual

Page 87

Advertising
background image

MONITORING

4

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

4-43

Troubleshooting Tips

Table 4-5 Troubleshooting Tips for NIBP Monitoring

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

NIBP AIR LEAK

message appears

Cuff applied too loosely

Leak in cuff/monitor pneumatic
system

• Check cuff for snug fit on

patient.

• Check that the cuff/monitor

connection is secure.

• Check cuff for leaks. Do not

use a cuff that exhibits a
leak.

NIBP FLOW ERROR

message

appears

The pneumatic system is not
maintaining stable cuff
pressure

• Deflate or remove cuff.
• Check tubing for leaks.
• Replace cuff.

NIBP FAILED

message appears

The monitor cannot establish
zero-pressure reference

• Check tubing for kink or

blockage.

• If this message persists,

remove monitor from use
and obtain service. Use
another method to measure
the patient’s blood pressure.

NIBP INITIALIZING

message

appears

NIBP requested while NIBP
module is still initializing

• Wait until message

disappears and request
NIBP.

NIBP MOTION

message appears

The patient extremity moved
too much for the monitor to
accurately complete the
measurement

• Have patient lie quietly with

extremity relaxed and
supported.

• Check that patient’s arm

does not move during NIBP
measurement.

NIBP OVERPRESSURE

message

appears

Cuff pressure exceeded
290 mmHg

• Disconnect tubing or remove

cuff.

• Avoid very rapid squeezing

of the cuff.

• If this message persists,

remove the cuff from use
and obtain service.