Batteries – Physio-Control LIFEPAK 15 User Manual

Page 34

Advertising
background image

3-16

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

Batteries

To install a battery:

1. Confirm that the battery is fully charged, unless the battery will be charged in the monitor/

defibrillator using the power adapter.

2. Inspect battery pins and contacts in the battery wells for signs of damage.

3. Align battery so battery clip is over the pins in the battery well.

4. Insert the end of the battery that is opposite the battery clip into the battery well.

5. Firmly press the clip end of the battery into the battery well until it clicks into place.

6. Repeat Step 1 through Step 5 to insert second battery.

To remove a battery, press the battery clip in and tilt the battery out of the battery well.

For information about battery maintenance, see "Battery Maintenance" on page 10-12.

WARNING

POSSIBLE LOSS OF POWER DURING PATIENT CARE

Battery pins in the defibrillator may be damaged if batteries are dropped or forced into

battery wells. Inspect pins routinely for signs of damage. Keep batteries installed at all

times except when the device is removed from service for storage.