– Skin (Note: Measurement Specialties skin temperature probe 4499HD is approved for use

with the LIFEPAK 15 monitor/defibrillator. Do not use Measurement Specialties part number

4499.)

For a list of the accessories that are intended for use with the LIFEPAK 15 monitor/defibrillator,

contact your Physio-Control representative or see the LIFEPAK 15 Monitor/Defibrillator Accessory

Catalog at www.physio-control.com. Carefully read the Instructions for Use that are provided with

the probes and connector cable for sensor placement instructions, use instructions, warnings,

cautions, and specifications.

IMPORTANT!

The Instructions for Use that are provided with the Measurement Specialties

temperature probes refer to a connector cable that is not compatible with the LIFEPAK 15 monitor/

defibrillator. Only use the adapter cable that is approved for use with the LIFEPAK 15 monitor/

defibrillator.

Temperature Monitoring Procedure

1. Connect the temperature adapter cable to the TEMP port on the monitor/defibrillator.

2. Connect the temperature probe to the temperature adapter cable.

3. Attach the temperature probe to the patient as described in the temperature probe Instructions

for Use.

Notes:

• The temperature area on the display is not activated until the monitor/defibrillator detects a

temperature between 24.8° and 45.2°C (76.6° and 113.4°F). To manually activate the

temperature monitoring area, use the speed dial to outline and select the temperature area

on the Home Screen. From the menu, select

ON

.

• The temperature probe may require 3 minutes to equilibrate after placement on the patient

monitoring site.

4. Confirm that the temperature reading appears and is stable.

5. Use the default label

TEMP

or select one of the user-selectable labels shown in Table 4-10. To

change the label, select the

TEMP

area. From the menu, select

TEMP

. Select a label from the list.

Table 4-10 TEMP Labels and Descriptions

LABEL

DESCRIPTION

T-esoph

Esophageal Temperature

T-naso

Nasopharangeal Temperature

T-bladder

Bladder Temperature

T-rectal

Rectal Temperature

T-skin

Skin Temperature