BASIC ORIENTATION

3

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

3-21

Note:

Older or heavily used batteries lose charge capacity. If a fully charged battery is installed in

the defibrillator and the battery status indicator shows less than four bars, the battery has reduced

capacity. If a battery status indicator shows only one or two bars after a fully charged battery is

installed, the battery has less than half the normal use time and should be recycled.

Alarms

LIFEPAK 15 monitor/defibrillator alarms can be set up to be ON or OFF when the defibrillator is

turned on. For more information, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided

with your device.

When alarms are set up to be ON, default limits are set. The limits temporarily appear to the right of

the active vital signs. For all vital sign default alarm limits, see Table A-3 on page A-14.

If alarms are set up to be OFF, press

ALARMS

to enable the alarms. Whether alarms are set up to be

ON or are enabled by pressing

ALARMS

, they can only be turned off by pressing

ON

to turn off the

device. If power is lost for less than 30 seconds, for example due to a system reset or changing the

only active battery, alarm settings are restored automatically.

Setting Alarms

When you press

ALARMS

, the following menu appears:

Select QUICK SET to activate the alarms for all
active monitoring functions.

The Quick Set limits automatically set high and
low limits based on the patient’s current vital
sign values. For example, if the patient’s HR is
70, selecting WIDE results in a high limit of
110 and a low limit of 45; selecting NARROW
results in a high limit of 100 and a low limit of
50. The default is WIDE.

Alarms

Quick Set

Limits

Silence

VF/VT Alarm

Wide

2 Min

Off

Select LIMITS to change alarm limits to WIDE or
NARROW

. See Table A-3 on page A-14.

Select SILENCE

to turn off the audible alarm for

up to 15 minutes. If an alarm limit is exceeded
while the alarm is silenced, the violated vital
sign flashes and an alarm message appears,
but the alarm tone remains silent.

Alarms

Quick Set

Limits

Silence

VF/VT Alarm

Wide

Narrow