Troubleshooting tips, Therapy – Physio-Control LIFEPAK 15 User Manual

Page 133

Advertising
background image

THERAPY

5

©2007-2013 Physio-Control, Inc.

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

5-19

Troubleshooting Tips

Table 5-1 Troubleshooting Tips for AED Mode

OBSERVATION

POSSIBLE CAUSE

CORRECTIVE ACTION

CONNECT ELECTRODES

message

appears

Therapy electrodes are not
connected to the therapy cable

• Check for electrode

connection.

Electrodes do not adhere
properly to the patient

• Press electrodes firmly on

patient’s skin.

• Clean, shave, and dry the

patient’s skin as
recommended.

• Replace the electrodes.

Electrodes are dry, damaged,
or out of date

• Apply new electrodes.

Therapy cable damaged

• Replace therapy cable and

perform daily checks per
Operator’s Checklist.

CONNECT CABLE

message

appears

Therapy cable is disconnected
during charging

• Reconnect cable and press

CHARGE

again.

Therapy cable damaged

• Replace therapy cable and

perform daily checks per
Operator’s Checklist.

MOTION DETECTED

and STOP

MOTION

messages appear during

analysis

Patient movement

• Stop CPR during analysis.
• When patient is being

manually ventilated, press
ANALYZE

after complete

exhalation.

Patient movement because of
agonal respirations

• Allow analysis to proceed to

completion—analysis is
delayed no more than
10 seconds due to motion
detection.

Electrical/radio frequency
interference

• Move hand-held

communication devices or
other suspected devices away
from the defibrillator, when
possible.

Vehicle motion

• Stop vehicle during analysis.
• Move patient to stable

location, when possible.

Advertising