10-8

LIFEPAK 15 Monitor/Defibrillator Operating Instructions

General Maintenance and Testing

11.Press

PRINT

.

12.Apply firm pressure with both paddles on the defibrillator checker paddle plates, and

simultaneously press and hold both

(shock) buttons while observing the screen.

13.Confirm that the defibrillator discharges on the next sensed QRS complex.

14.Press

PRINT

again to stop the printer.

15.Confirm that the defibrillator returns to Asynchronous mode (sense markers are no longer

displayed and

SYNC

LED is off).

Note:

Defibrillator may be set up to remain in Sync mode after discharge.

16.Confirm that the printer annotates the time, date, Sync On, sense markers prior to energy

delivered, energy selected, no sense markers after Shock 1, and Sync Off on the ECG strip.

17.Turn off the defibrillator.

Note:

If a

CONNECT CABLE

,

PADDLES LEADS OFF

, or any other warning message appears, replace the

paddle assembly with a new paddle assembly and repeat the test. If the problem cannot be

corrected, remove the device from active use and contact a qualified representative.

WARNING

POSSIBLE PADDLE DAMAGE AND PATIENT BURNS

Press paddles firmly onto the defibrillator checker plates when discharging to prevent

arcing and formation of pits on paddle surfaces. Pitted or damaged paddles may cause

patient skin burns during defibrillation.